Manual programming:
The manual mode permits to set voltage, current,
power and time for a routine electrophoresis run. Parameters can be
changed temporarily without interrupting the run.
Method programming:
Up to 9 different programs, each with 9 steps, of
frequently used parameters can be stored in the non-volatile memory for
future recall. Reads voltage in 1 V steps, current in 1 mA steps and
power in 1 W steps. Each step is able to recall a next one, providing a
flexible multiple step function for special techniques. Parameters of
the running step can be changed temporarily without interrupting the
run.
Voltage ramp:
The method mode also permits to program a linear
voltage gradient for any step provided the limiting current or power is
not attained.
Timer:
Timer or volt-hour controlled operation is a
useful standard feature on all models. The microcomputer will
automatically terminate the run when the count down of the selected
value is achieved.
Automatic cross-over:
Each model has constant voltage, constant current,
constant power capabilities with automatic cross-over and shows which
parameter is kept constant.
Automatic recovery
after power failure:
In case of a mains drop out the instrument will
automatically continue the run for the remaining time.
Data-logging:
Stores periodically up to 3600 output values
(voltage, current and power, time or volthours) including program number
and step.
Data Transfer:
A free data acquisition software for PC is
available optionally.
Download software
(attention:
.NET Framework 1.1 is required!). It
permits to visualise and examine the stored run details.
Remote control:
All power supplies can be controlled by a computer using special
commands.
Safety precautions:
The user is protected from potential shock hazard
since the AC line is automatically disconnected from the high voltage
transformer when a ground leakage path is detected. The instrument is
fully protected against any overload condition including accidental
short circuit of the output. The high voltage cannot suddenly appear at
the outputs. It will always increase smoothly up to one of the pre-set
limits is reached. Galvanic RS232 input/output insulation prevents
dangerous ground loop interferences when connected to a computer.
(Model EV222 has no method programming,
data-logging or remote control possibility)
|

Models:
EV222:
0...200 V, 0...200 mA, 0...20 W
EV243:
0...400 V, 0...300 mA, 0...50 W
EV231:
0...300 V, 0...1000 mA, 0...150 W
EV265:
0...600 V, 0...500 mA, 0...150 W
EV202:
0...300 V, 0...2000 mA, 0...300 W
EV261:
0...600 V, 0...1000 mA, 0...300 W
EV215:
0...1200 V, 0...500 mA, 0...300 W
EV232:
0...3000 V, 0...150 mA, 0...150 W
EV233:
0...3000 V, 0...300 mA, p0...300 W
EV262:
0...6000 V, 0...150 mA, 0...300 W |